NetScanner™ System (9016, 9021, & 9022) User’s Manual

NOTE: Selecting too many other data groups will compromise module performance.

bbbb(hex) data selected for inclusion in each stream packet

0001 ** Enable Valve Position Status (reserved for future use)

0002 Enable DH Temperature Status (see bit map below)

0010 Enable Pressure EU Data (default if “05” never executed after “00”)

0020 Enable Pressure A/D Counts

0040 Enable Pressure Voltages

0080 Enable DH Temperature EU Data (degrees C)

0100 Enable DH Temperature A/D Counts

0200 Enable DH Temperature Voltages

Any DH Temperature Status datum is delivered as a two-byte binary bit map (16- bit, big endian) with each bit representing the status of DH #16 through DH #1 respectively. A bit value of 0 (zero) indicates the DH is operating within its specified limits. A value of 1 (one) indicates the DH is outside its specified limits.

Bit #

15

14

13

12

11

10

9

8

7

6

5

4

3

2

1

0

Chan #

16

15

14

13

12

11

10

9

8

7

6

5

4

3

2

1

Binary

1

0

0

0

0

0

0

0

0

0

0

0

0

0

0

1

Hex

 

8

 

 

 

 

0

 

 

0

 

 

 

 

1

 

The above example indicates that Channels 1 and 16 are operating outside the

specified temperature limits.

**NOTE: This status field (0001) cannot be specified for Models 9016, 9021, or 9022. However it is shown should the capability be added to future firmware versions. Currently, only Models 9816 and 903x can return Valve Position status in their streams.

3 - 44www.PressureSystems.com