802BT/802R Video Test Generator User Guide 713
-OR-
The generator hardware configuration does not support a video format with a total (active
+ blanked) number of pixel clock cycles per horizontal scan line that is not evenly divisible
by 2 for the selected analog or digital composite sync type.
2091 Horizontal total not a multiple of 4
The generator hardware configuration does not support a video format with a total (active
+ blanked) number of pixel clock cycles per horizontal scan line that is not evenly divisible
by 4. Mostly applies to video formats using serial digital video.
2092 Horizontal resolution not even
Number of active pixels per line must be evenly divisible by 2 for the given video format
type and generator hardware configuration.
2093 Horizontal sync pulse width not even
Number of pixel clock cycles in the horizontal sync pulse width must be evenly divisible by
2 for the given video format type and generator hardware configuration.
2094 Horizontal sync pulse delay not even
The number of pixel clock cycles in the horizontal sync pulse delay must be evenly
divisible by 2 for the given video format type and generator hardware configuration.
2096 Horizontal resolution not a multiple of 4
The generator hardware configuration does not support a video format with a number of
active pixels per horizontal scan line that is not evenly divisible by 4. Mostly applies to
video formats using serial digital video.
2097 Horizontal sync pulse width not a multiple of 4
The generator hardware configuration does not support a video format with the number of
pixel clock cycles in the horizontal sync pulse width not evenly divisible by 4. Mostly
applies to video formats using serial digital video.
2098 Horizontal sync pulse delay not a multiple of 4
The generator hardware configuration does not support a video format with the number of
pixel clock cycles in the horizontal sync pulse delay not evenly divisible by 4. Mostly
applies to video formats using serial digital video.
2099 JRAT greater than 200MHz not allowed
The 802BT generator does not support a setting of JRAT greater than 200MHz.