Sigma SQM-160 manual Sensor Selection, Chapter, Operation

Models: SQM-160

1 35
Download 35 pages 41.43 Kb
Page 18
Image 18
2.4Sensor Selection

Chapter 2

Operation

2.4Sensor Selection

The SQM-160 comes standard with two sensor inputs. Four additional sensors are available by adding a Sensor Option Card. The sensors can each be assigned to a different film, or multiple sensors can be averaged for a single film. The averaging option provides more uniform coverage of the deposition area, and provides a backup sensor capability. If one of multiple sensors assigned to a film fails, the sensor is automatically removed from rate/thickness calculations.

To assign a sensor, or sensors, to a film:

1.Press ProgramSensor status can be seen by observing the Crystal Status LEDs: If the LED is not illuminated, the crystal is disabled. to enter program mode.

2.Use the Control Knob to scroll to the desired Film # (1-9).

3.Depress the Control Knob or NextManual background to enter the film parameters menu for the selected film.

4.Press NextManual background until SENS AVG is shown.

5.Use the Control Knob to scroll through the sensors in Display2.

6.Depress the Control Knob to toggle the sensor on/off.

Sensor status can be seen by observing the Crystal Status LEDs: If the LED is not illuminated, the crystal is disabled.

If the LED is illuminated,the crystal is enabled and receiving valid readings.

If the LED is flashing, the crystal is enabled, but is not receiving valid readings.

7.Continue selecting sensors until the Crystal Status LEDs indicate the desired setup.

8.Press ProgramManual background to exit the Film Parameters Menu and return to normal mode.

2-7

Page 18
Image 18
Sigma SQM-160 manual Sensor Selection, Chapter, Operation