Sigma SQM-160 manual Four Sensor Card, Analog Output Card, Chapter, Options

Models: SQM-160

1 35
Download 35 pages 41.43 Kb
Page 24
Image 24
3.2Four Sensor Card

Chapter 3

Options

3.2Four Sensor Card

The SQM-160 automatically detects the presence of the Four Sensor Card. Refer to System Connections in Chapter 1 for sensor hookup information.

Refer to Chapter 2, Sensor Selection, to assign Four Sensor Card sensors to a film.

3.3Analog Output Card

Once the Analog Output Card is installed, the SQM-160 System Parameters must be set to match the device that will be attached to the Rate or Thickness output.

To setup the Analog Output Card in the System Parameters menu:

1.Press Program3.3Analog Output Card to enter program mode.

2.Press PrevManual background to enter the System Parameters Menu

3.Use Next Manual background to move through the system parameters until R/T BNDS is displayed.

4.Depress the Control Knob to display RATE MIN.

5.Adjust the Control Knob to the Rate desired for a 0V output.

6.Depress the Control Knob to save the value and display the RATE MAX setting.

7.Adjust the Control Knob to the Rate desired for a 5V output.

8.Repeat steps 4-7 to adjust the Thickness output values.

9.Press ProgramManual background to exit the System Parameters Menu and return to normal mode.

Refer to System Parameters Setup in Chapter 2 for more information on setting SQM- 160 System Parameters.

3-2