Sigma SQM-160 System Parameters Menu, Chapter, Operation, Sub-Menu, Description, Range, Default

Models: SQM-160

1 35
Download 35 pages 41.43 Kb
Page 16
Image 16
System Parameters Menu

Chapter 2

Operation

System Parameters Menu

Sub-Menu

Program Prev

-TIMEBASE

-SIM MODE

-FRQ DISP

-RATE RES

-RATEFILT

-BAUDRATE

-ETCH

-xTOOLING

-FMIN/MAX

-R/T BNDS

-xTLTOOL1

-xTLTOOL2

-xTLTOOL3

.

.

-xTLTOOL6

-FREQ MIN

-FREQ MAX

-RATE MIN

-RATE MAX

-THICKMIN

Display

TIMEBASE

SIM MODE FRQ DISP

RATE RES RATEFILT

BAUDRATE ETCH

Description

Range

Default

Units

Time required to take a single

0.15 – 2.00

0.25

sec

measurement

 

 

 

Simulates sensors.

On/Off

Off

 

Toggles between Rate/Thickness

On/Off

Off

 

and Frequency Displays

 

 

 

Display resolution of Rate

Hi/Low

Low

 

Number of readings averaged for

1 – 20

8

 

rate calculations

 

 

 

Serial Baud Rate to PC

2.4 – 19.2

19.2

kbps

Sets mode to negative rate for

On/Off

Off

 

etching applications

 

 

 

2-5

Page 16
Image 16
Sigma SQM-160 System Parameters Menu, Display TIMEBASE SIM MODE FRQ DISP RATE RES RATEFILT BAUDRATE ETCH, Chapter, Range