Film Selection Film Parameters Menu

Chapter 2

Operation

The diagram and table below detail the parameters available in the Film Parameters Menu:

Film Selection Film Parameters Menu

Sub-Menu

Program

Film1

.

.

.

.

.

.

Film9

- DENSITY

- TOOLING

- Z-FACTOR

- FINL THK

- THK SET

- TIME SET

- SENS AVG

-1

-2

- 3

Sensors 3 to 6 show only if the

. four sensor

. option card is - 6 iinstalled.

Display

DENSITY

TOOLING

Description

Range

Default

Units

Density of the material being

0.5 – 99.99

1.00

gm/cc

deposited.

 

 

 

Overall tooling Factor for this film.

10 – 399

100

%

See the Sensor Tooling section of

 

 

 

this chapter for information on setting tooling values.

Z-FACTOR

FINL THK

Z-Factor of the material being

0.10 – 10.00

1.0

 

deposited.

 

 

 

Desired Final Thickness of

0.000 – 99.99

0.500

deposited material. Lights Final Thk

 

 

 

LED. Zero is disabled.

 

 

 

THK SET Thickness value that closes the 0.000 – 99.99 0KÅ Thickness Setpoint relay and lights

Thk SP LED. Zero is disabled.

TIME SET

SENS AVG

Elapsed time that closes the Timer

0:00 – 99:59

0

Sec

Setpoint relay and lights Time SP

 

 

 

LED. Zero is disabled.

 

 

 

Enable/disable crystals for this film.

Enabled/

Ch1

 

See the Sensor Selection section of

Disabled

Enabled

 

this chapter for more information.

 

 

 