M4557 - Model 9210 Nitriding Controller

The next screen is the Zero Input 3 Range 2 screen. This screen will allow the user to set the zero scale for Input 3 Range 2. A value of 0 millivolts will need to be sourced to the inputs. For a zero calibration, enter a 0 as the value of the terminal to correctly calibrate the inputs. The current Input 3 value will be displayed near the bottom of the screen as “Current Input 3 value: xxxx.xx uV”. The progress bar at the bottom of the screen will display the calibration progress.

The next screen is the Span Input 3 Range 2 screen. This screen will allow the user to set the span value for Input 3 Range 2. A suggested value will be displayed next to the Calibrate button (“sugg. 65.000 mV”). The current Input 3 value will be displayed near the bottom of the screen as “Current Input 3 value: xxxx.xx uV”. The progress bar at the bottom of the screen will display the calibration progress.

The next screen is the Zero Output 1 screen. This screen will allow the user to set the zero value for Output 1. Measured at terminals 24(-) and 25(+) for this step. The progress bar at the bottom of the screen will display the calibration progress.

Zero Output 1

Enter span output current (mA)

Calibrate

Edit

<-- Back

 

Skip

 

Next -->

 

DONE

Span Output 1

Entered measured output current (mA) Calibrate

Edit

<-- Back

 

Skip

 

Next -->

 

DONE

The next screen is the Zero Output 2 screen. This screen will allow the user to set the zero value for Output 2. Measured at terminals 26(-) and 25(+) for this step. The progress bar at the bottom of the screen will display the calibration progress.

The next screen is the Span Output 1 screen. This screen will allow the user to set the span value for Output 1. Measured at terminals 24(-) and 25(+) for this step. The progress bar at the bottom of the screen will display the calibration progress.

Zero Output 2

Enter zero output current (mA)

Calibrate

Edit

<-- Back

 

Skip

 

Next -->

 

DONE