M4557 - Model 9210 Nitriding Controller
The next screen is the Zero Input 3 Range 2 screen. This screen will allow the user to set the zero scale for Input 3 Range 2. A value of 0 millivolts will need to be sourced to the inputs. For a zero calibration, enter a 0 as the value of the terminal to correctly calibrate the inputs. The current Input 3 value will be displayed near the bottom of the screen as “Current Input 3 value: xxxx.xx uV”. The progress bar at the bottom of the screen will display the calibration progress.
The next screen is the Span Input 3 Range 2 screen. This screen will allow the user to set the span value for Input 3 Range 2. A suggested value will be displayed next to the Calibrate button (“sugg. 65.000 mV”). The current Input 3 value will be displayed near the bottom of the screen as “Current Input 3 value: xxxx.xx uV”. The progress bar at the bottom of the screen will display the calibration progress.
The next screen is the Zero Output 1 screen. This screen will allow the user to set the zero value for Output 1. Measured at terminals
Zero Output 1
Enter span output current (mA) | Calibrate |
Edit
| Skip |
| Next |
| DONE |
Span Output 1
Entered measured output current (mA) Calibrate
Edit
| Skip |
| Next |
| DONE |
The next screen is the Zero Output 2 screen. This screen will allow the user to set the zero value for Output 2. Measured at terminals
The next screen is the Span Output 1 screen. This screen will allow the user to set the span value for Output 1. Measured at terminals
Zero Output 2
Enter zero output current (mA) | Calibrate |
Edit
| Skip |
| Next |
| DONE |