M4557 - Model 9210 Nitriding Controller

 

 

 

 

 

 

System errors:

 

 

 

 

 

 

 

Code

Error

Description

 

 

 

 

 

 

 

71-76

NVRAM CRC error

Three areas exist (1,3 & 5) each with a backup (2,4 & 6). Area 1

 

 

 

 

contains user parameters i.e. calibration data and the backup is write

 

 

 

 

protected so if this area is restored user calibration etc is LOST.

 

 

 

 

Error 71 will be reported in response to any read request until either

 

 

 

 

a calibration is performed or the instrument is rebooted. Areas 3 & 5

 

 

 

 

are read only so they can be silently restored. Errors 73 & 75

 

 

 

 

therefore should never be reported, although they exist internally

 

 

 

 

until they clear themselves. Errors 72,74 & 76 will be reported in

 

 

 

 

response to any read request and cannot be cleared.

 

 

 

 

 

 

 

77- 78

TCD curve error

In reading curve data from EEPROM the instrument has attempted

 

 

 

 

to read beyond the ends of the curve.

 

 

 

 

 

 

 

79

Wrong block no.

The serial numbers of the Nitriding Sample Cell stored in areas 1 & 5

 

 

 

 

(in separate devices) do not match

 

 

 

 

 

 

 

80

UART missing/error.

For debugging only. Since the only interface available is RS232, this

 

 

 

 

error can never be reported!

 

 

 

 

 

 

 

81

Reserved

For external EEPROM not found.

 

 

 

 

 

 

SSi Manual SERIES 9210-M4557-Nitriding

Page 62 of 62

Page 62
Image 62
SSI America 9210 user manual System errors