Acquiring Waveforms
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual 3-57
FastFrame is not compatible with these features or modes:
HEquivalent Time
HHistograms
HFast Acquisitions
HAverage
HEnvelope
HWaveform Database
Consider the following operating characteristics when using FastFrame:
HYou can push RUN/STOP to terminate a FastFrame sequence. If any frames
were acquired, they are displayed. If no frames were acquired, the previous
FastFrame waveform is displayed.
HBecause FastFrame introduces additional processing time into the operation
cycle of acquire, process, and display, its best to use Single Sequence
Acquisition (see Acquire menu, Stop After menu). With Single Sequence
selected, you will see the current acquisition sequence; otherwise, the display
lags the current sequence by one sequence. You can also see the current
sequence by pushing the RUN/STOP button to stop the acquisition.

Using FastFrame

Acquisitions