Displaying Waveforms
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual 3- 143
Overview Related control elements and resourcesSet display styles (Cont.)
Select a
persistence
mode
6. From the the Display setup control window (see right),
choose a persistence mode:
HInfinite Persistence to make data persist indefinite-
ly. Waveform displays accumulate data as new
waveform records acquire, resulting in a build up of
data in the displayed waveforms.
HVariable Persistence to make data persist
momentarily, but also decay. New waveform
displays accumulate data as new waveform records
acquire, but with continuous replacement of the
oldest data.
If you select Variable Persistence, set a time at
which the oldest data fades away.
HOff to make data display for the current acquisition
only.
HReset to restart the accumulation of data.
Continue with
next
procedure
7. For more ways to customize the display, see the next
procedure.
See Customize Graticule and Waveforms on page 3--144.