Incoming Inspection
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual 1-35

Perform the Extended Diagnostics

Extended diagnostics and self calibration perform a more detailed functionality
check than the incoming inspection and Power-on diagnostics.
NOTE. Allow a 20-minute warm-up before running the self calibration.
Disconnect any attached probes from the instrument. Then select the Utilities
menu. Run the self calibration followed by the extended diagnostics by first
selecting the Instrument Calibration (see page 3--171 for additional information
on signal path compensation) and then the Instrument Diagnostics tabs. Results
of the tests display on their property pages.
To check the hardware and Windows software underlying the instrument UI (user
interface), run the CheckIt Utilities from the Windows Start menu:
1. Exit the instrument application before running the external diagnostics.
a. Display the Task Manager by pressing CTRL,ALT,andDelete,and
then touch Task Manager.
b. Select TekScope and then touch End Task.
2. Click Start, then click Programs in the Start Menu. Finally, click CheckIt
Utilities.
3. From the Go To menu, select and run the tests you want to perform.
4. Check test results. All tests except the Modem and CD-ROM tests should
pass. The CD-ROM test requires data from the CheckIt Utilities CD.
5. Dismiss the CheckIt Utilities: Select Exit in the File menu.
6. Restart your instrument UI software: Click Start, then click Programs in the
Start Menu. Finally, click Tektronix Oscilloscope and then TekScope.

Checking the Underlying

System (Optional)