Appendix A: Specifications
A-40 CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Table A- 12: Certifications and compliances (Cont.)
Category Standards or description
EC Declaration of Conformity -- EN 61000-3-2 Power Harmonic Current Emissions
y
EMC 1If interconnect cables are used, they must be low-EMI shielded cables such as the following
Tektronix part numbers or their equivalents: 012-0991-01, 012-0991-02 or 012-0991-03 GPIB
Cable; 012-1213-00 (or CA part number 0294-9) RS-232 Cable; 012-1214-00 Centronics
Cable; or LCOM part number CTL3VGAMM-5 VGA Cable.
2CSA7404, CSA7154, TDS7404, TDS7254, TDS7154, TDS6604, & TDS6404:
The performance criteria for when the instrument is subjected to the conditions described
above are defined as follows:
A—10 mV/division to 1 V/division: ≤0.4 division waveform displacement or ≤0.8 division
increase in peak-to-peak noise
5 mV/division and 2 mV/division, typical: ≤8 mV increase in peak-to-peak noise
B—temporary, self-recoverable degradation or loss of performance is allowed, but no change
of actual operating state or loss of stored data is allowed
C—temporary loss of function is allowed provided that the function is self recoverable or can
be restored by the operation of the controls
TDS7104 & TDS7054:
The performance criteria for when the oscilloscope is subjected to the conditions described
above are defined as follows:
A—≤0.2 division waveform displacement or ≤0.4 division increase in peak-to-peak noise
B—temporary, self-recoverable degradation or loss of performance is allowed, but no change
of actual operating state or loss of stored data is allowed
C—temporary loss of function is allowed provided that the function is self recoverable or can
be restored by the operation of the controls
3Radiated emissions may exceed the levels specified in EN 61326 when this instrument is
connected to a test object.
4USB mouse and keyboard only, performance criterion C. Normal USB keyboard or mouse
operation can be restored by unplugging and reconnecting the USB connector to the
instrument.
Australia/New Zealand Declara-
tion of Conformity —EMC
CSA7404, CSA7154, TDS7404,
TDS7254, TDS7154,
TDS6604, & TDS6404
Complies with EMC provision of Radiocommunications Act per the following standard(s):
AS/NZS 2064.1/2 Industrial, Scientific, and Medical Equipment: 1992, Class A
FCC
TDS7104 & TDS7054:
Radiated and conducted emissions do not exceed the levels specified in FCC47 CFR, Part 15,
Subpart B, for Class A equipment.