CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual A-1

Appendix A: Specifications

This chapter contains the specifications for the CSA7000 Series Communica-
tions Signal Analyzers, the TDS7000 series Digital Phosphor Oscilloscopes, and
TDS6000 Series Digital Storage Oscilloscopes. All specifications are guaranteed
unless labeled typical.Typical specifications are provided for your conve-
nience but are not guaranteed. Specifications that are marked with the nsymbol
are checked in chapter four, Performance Verification, of the service manual.
All specifications apply to all models unless noted otherwise. To meet specifica-
tions, the following conditions must be met:
HThe instrument must have been calibrated in an ambient temperature
between 20 °C and 30 °C(68°F and 86 °F).
HThe instrument must be operating within the environmental limits listed in
Table A--11 on page A--38.
HThe instrument must be powered from a source that meets the specifications
listed in Table A--9 on page A--37.
HThe instrument must have been operating continuously for at least 20 min-
utes within the specified operating temperature range.
HYou must perform the Signal Path Compensation procedure after the
20-minute warm-up period, and if the ambient temperature changes more
than 5 °C(41°F), repeat the procedure. See Optimizing Measurement
Accuracy on page 3--171 for instructions to perform this procedure.