Incoming Inspection
1-30 CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
d. Adjust the horizontal delay: Rotate the upper multipurpose knob to
change the horizontal delay setting. Verify that the falling edge shifts
horizontally. Rotate the front-panel horizontal POSITION knob. Verify
that this knob has the same effect (it also adjusts delay, but only when
delay mode is on).
e. Verify the delay toggle function:
HRotate the front-panel horizontal POSITION knob to center the
falling edge horizontally on the screen.
HChange the horizontal SCALE to 40 ns/div. The falling edge of the
waveform should remain near the center graticule. If not, readjust the
delay setting to center the falling edge.
HPush the front-panel DELAY button several times to toggle delay off
and on and back off again. Verify that the display switches quickly
between two different points in time (the rising and falling edges of
this signal).
9. Remove the test hookup: Disconnect the equipment from the channel input
and the probe compensation output.
Equipment
required
One BNC cable
CSA7404, CSA7154, TDS7404, TDS7254, TDS7154,TDS6604, &
TDS6404
One TCA-BNC TekConnect adapter
Prerequisites None
1. Initialize the instrument: Push the front-panel DEFAULT SETUP button.
2. Hook up the signal source: Connect the equipment to the CH 1 input as
shown in Figure 1--10.
3. Set up the instrument: Push the front-panel AUTOSET button.

Check Trigger Operation