Appendix A: Specifications

CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual A-15

Table A- 2: Channel input and vertical specifications (Cont.)
Characteristic Description
nOffset accuracy Net offset is the nominal voltage that must be applied to the channel to bring the trace
to center screen. Net offset = offset -- ( position ×volts/division). Offset accuracy is
the accuracy of this voltage level.
CSA7404, CSA7154, TDS7404,
T
D
S
7
2
5
4
&
T
D
S
7
1
5
4
SCALE range Offset accuracy
TDS7254, & TDS7154 2mV/divto9.95mV/div ±(0.2% ×| net offset | + 1.5 mV + 0.1
div ×V/div setting)
10 mV/div to 99.5 mV/div ±(0.35% ×| net offset | + 1.5 mV + 0.1
div ×V/div setting)
100 mV/div to 1 V/div ±(0.35% ×| net offset | + 15 mV + 0.1
div ×V/div setting)
TDS7104 & TDS7054 SCALE range Offset accuracy
1 mV/div to 100 mV/div ±(0.2% ×| net offset | + 1.5 mV +
0.1 div ×V/div setting)
101 mV/div to 1 V/div ±(0.25% ×| net offset | + 15 mV +
0.1 div ×V/div setting)
1.01 V/div to 10 V/div ±(0.25% ×| net offset | + 150 mV +
0.1 div ×V/div setting)
TDS6604 & TDS6404 SCALE range Offset accuracy
10 mV/div to 99.5 mV/div ±(0.5% ×| net offset | + 1.5 mV +
0.1 div ×V/div setting)
100 mV/div to 1 V/div ±(0.5% ×| net offset | + 15 mV + 0.1 div
×V/div setting)