Appendix A: Specifications

A-28 CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual

Table A- 4: Trigger specifications (Cont.)
Characteristic Description
Trigger level or threshold accuracy, typical
CSA7404, CSA7154, TDS7404,
TDS7254, & TDS7154
Edge trigger, DC coupling, for signals having rise and fall times 1ns
Trigger Source Accuracy
Any channel ±[(2% ×| setting -- net offset | ) +
(0.35 div ×volts/div setting) + offset
accuracy] (10 mV/div only)
Auxiliary Not specified
where, net offset = offset -- ( position ×volts/division)
TDS7104 & TDS7054 Edge trigger, DC coupling, for signals having rise and fall times 20 ns
Trigger Source Accuracy
Any channel ±[(2% ×| setting -- net offset | ) +
(0.3 div ×volts/div setting) + offset
accuracy]
Auxiliary Not calibrated or specified
where, net offset = offset -- ( position ×volts/division)
TDS6604, & TDS6404 Edge trigger, DC coupling, for signals having rise and fall times >1 ns
Trigger Source Accuracy
Any channel ±[(2% ×| setting -- net offset | ) +
(0.35 div ×volts/div setting) + offset
accuracy] (10 mV/div only)
Auxiliary Not specified
where, net offset = offset -- ( position ×volts/division)
B Event (Delayed) trigger Trigger After Time Trigger on nth Event
Range Delay time = 16 ns to 250 s Event count = 1 to 107
Minimum time between arm (A Event)
and trigger (B Event), typical
2 ns from the end of the time period to
the B trigger event
2 ns between the A trigger event and the
first B trigger event
Minimum pulse width, typical B event width 1ns
Maximum frequency, typical B event frequency 500 MHz