2.SPECIFICATIONS AND AGENCY APPROVALS

2.1.Specifications

Measuring principle

UV absorption (Beer Lambert Law)

 

 

Ranges

User selectable to any full scale range from:

 

0-1000 PPM to 0-10,000 PPM

 

0-5 g/Nm3 to 0-50 g/Nm3

 

0-1 wt% to 0-5 wt%

Measurement Units

PPM, wt%, g/Nm3

Accuracy

1% of Full Scale

Precision/Repeatability

0.5% of Full Scale

Display Resolution

1 PPM, 0.001 wt%, 0.01 g/Nm3

Response Time (95%)

<45 seconds to 95%

Compensation

Pressure, Temperature (NTP = 273.15K, 760

mmHg)

 

Gas Inlet Pressure Range

11.0 – 16.0 psia

Sample Flow Rate

0.8 LPM

Temperature Range

5-45oC

 

Rack Mount (3U Panel) - 5.22” (133 mm) x 19.0"

Dimensions (H x W x D)

(483 mm) x 15.3" (388 mm)

 

With NEMA 4X enclosure – 16.01” (407 mm) x

 

15.77” (401 mm) x 8.02” (204 mm)

Weight

Rack Mount – 13.6 lbs (6.17 kg)

 

With NEMA 4X enclosure– 16.5 lbs (7.5 kg)

Power

100-240 VAC, 47-63 Hz, 74 W Max

Maximum Operating Altitude

2000 meters

Analog Output

0 – 5 V or 4-20 mA isolated output

Status Outputs

System OK, Invalid Reading, UV Lamp Fault,

Pneumatic Error

 

RS-232 (optional RS-485)

57.6 Kbaud, DB-9 connector