4. GETTING STARTED
The M465M is shipped with the following standard equipment:
1.Power cord (Rack mount version only)
2.Instruction manual.
Upon receiving the M465M please do the following:
1.Verify that there is no apparent shipping damage. (If damage has occurred please advise shipper first, then Teledyne API.)
4.1.Mechanical Installation (for NEMA 4X Enclosure)
Mount the enclosure securely to a vertical surface.
Figure
All four mounting holes should be used to secure the monitor.
Use stainless steel, 5/16” diameter bolts.
4.1.1.Ventilation Clearance
When installing the M465M be sure to leave sufficient ventilation clearance.
Table
Area | Minimum required clearance |
Back of the monitor | None |
Sides of the monitor | 1 inch |
Above and below the monitor | 1 inch |