Figure 7-4: Span Cal Menu
7.2.4.VIEW Menu
The VIEW menu allows the operator to view various measurement parameters in real- time. This can be useful for diagnosing various instrument or system problems.
NOTE
These values are updated in
Table
| Parameter |
| Meaning |
| Units |
| Normal Range* |
|
|
|
|
|
|
|
|
| MEAS |
| UV detector reading, measure cycle |
| mV |
| 250 – 1230 |
|
|
|
|
|
|
|
|
| REF |
| UV detector reading, reference |
| mV |
| 250 – 1230 |
|
| cycle |
|
| |||
|
|
|
|
|
|
| |
| PRESSURE |
| Sample pressure |
| psia |
| 9.0 – 18.0 |
|
|
|
|
|
|
|
|
| LAMP |
| UV lamp temperature |
| ºC |
| 50.0 – 52.0 |
| TEMP |
|
|
| |||
|
|
|
|
|
|
| |
| SAMPLE |
| Sample temperature |
| ºC |
| 5.0 |
| TEMP |
|
|
| |||
|
|
|
|
|
|
| |
| FLOW |
| Sample Gas Flow Rate** | ON/OFF |
| ON | |
|
|
|
|
|
|
| |
| *After initial 20 minute |
|
|
|
| ||
| ** Optional Hardware. |
|
|
|
|
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 39