Figure 7-4: Span Cal Menu

7.2.4.VIEW Menu

The VIEW menu allows the operator to view various measurement parameters in real- time. This can be useful for diagnosing various instrument or system problems.

NOTE

These values are updated in real-time as they are repeatedly scrolled on the display.

Table 7-5: VIEW Menu Parameters

 

Parameter

 

Meaning

 

Units

 

Normal Range*

 

 

 

 

 

 

 

 

 

MEAS

 

UV detector reading, measure cycle

 

mV

 

250 – 1230

 

 

 

 

 

 

 

 

 

REF

 

UV detector reading, reference

 

mV

 

250 – 1230

 

 

cycle

 

 

 

 

 

 

 

 

 

 

PRESSURE

 

Sample pressure

 

psia

 

9.0 – 18.0

 

 

 

 

 

 

 

 

 

LAMP

 

UV lamp temperature

 

ºC

 

50.0 – 52.0

 

TEMP

 

 

 

 

 

 

 

 

 

 

 

SAMPLE

 

Sample temperature

 

ºC

 

5.0 –50.0

 

TEMP

 

 

 

 

 

 

 

 

 

 

 

FLOW

 

Sample Gas Flow Rate**

ON/OFF

 

ON

 

 

 

 

 

 

 

 

*After initial 20 minute warm-up period.

 

 

 

 

 

** Optional Hardware.

 

 

 

 