10.2.1.Sensor OK
The Sensor OK LED indicates the status of the O3 sensor module in the monitor. The normal state of this LED (or Status Output) is On. If this LED remains off after the normal
The most common cause of this warning is a failure of the UV Lamp. A UV Lamp Adjustment (See Section 9.4) should be attempted. If the UV Lamp cannot be adjusted to meet the specified values, then a UV Lamp Replacement should be performed (See Section 9.5.)
10.2.2.Invalid Reading
The Invalid Reading LED indicates that the instrument is reading a value that cannot be represented properly on the analog output. Since the analog output is limited to 0- 5V (or
If the monitor is consistently reading negative values, then a zero calibration should be performed, see Section 7.2.3.4.
If the monitor is consistently reading values in excess of the
10.2.3.Check Lamp
The Check Lamp LED indicates that the UV Lamp intensity has dropped below 250mV, a level where UV Lamp Adjustment (See Section 9.4) should be made at the next convenient opportunity. Note that this is a
10.2.4.Pneumatic Error
The Pneumatic Error LED indicates that one of the pneumatic parameters, flow or pressure, has gone outside of normal ranges. Note that this is a
The first step in troubleshooting a Pneumatic Error is determining which parameter has caused the warning. At the monitor front panel, navigate to the VIEW menu (See Section 7.2.4.) Examine the Flow and Pressure values and compare them to the limits described in Table
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 67