7.2.6.SETUP-ALARM Menu
The ALARM menu allows the operator to view and edit properties of the concentration alarms. The table below summarizes the alarm configuration settings.
Table
Property | Meaning | Default | Range |
|
| Value |
|
HI ALARM | HI alarm trigger value | 10 PPM | 10 - 9999 |
|
|
| PPM |
HIHI ALARM | 20 PPM | 10 - 9999 | |
|
|
| PPM |
ON/OFF HI | HI alarm enable, enables HI alarm | ON | ON/OFF |
ALARM | operation |
|
|
ON/OFF HI HI | ON | ON/OFF | |
ALARM | alarm operation |
|
|
7.2.7.CONFIG Menu
The CONFIG menu displays important information about the instrument. The Serial Number, COMM address settings, and the latest revision of software installed are information available in the CONFIG menu. Settings can be changed under other menus (see Sections 7.2.3, 7.2.5, and 7.2.6); however the CONFIG menu just contains
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 41