10.2.Troubleshooting Using Front Panel Status LED’s or Status Outputs

The M465M has 4 front-panel status LED’s that reflect the current operating status of the monitor, and indicate fault conditions. There are also four relay Status Outputs on the rear panel (or internally in the NEMA configuration) that also reflect the state of these status LED’s. See Section 4.5.3 for more information on the Status Outputs.

Table 10-1: Status LED/Output Definitions

Status

Status LED

Normal

Triggers

Critical

Output #

Label

Operating

 

 

Warning?

 

 

State

 

 

 

1

Sensor OK

On

Reference < 125.0 mV

Yes

 

 

 

Reference > 1230.0 mV – 0.5 Wt%

 

2

Invalid Reading

Off

O3 Concentration < F.S. Range

No

 

 

 

O3 Concentration > F.S. Range

 

3

Check Lamp

Off

Reference < 250.0 mV

No

4

Pneumatic

Off

Pressure < 9.0 psia

No

 

Error

 

Pressure > 18.0 psia

 

 

 

 

Flow < 500 cc/min*

 

*If Flow

Switch installed

 

 

 

 

P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 66

Page 68
Image 68
Teledyne 465M instruction manual Status LED/Output Definitions