P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 40
7.2.5. SETUP-VARS Menu
The VARS menu allows viewing and editing of various global setup variables that effect
how the M465M operates. These variables are stored in the instrument’s non-volatile
memory. Table 7-6 lists these variables and their function.
Table 7-6: VARS List
VAR Meaning Default
Value Range
ALARM MODE
Controls the latching behavior of
the concentration alarms. See
Section 7.1.3 for additional
details.
LATCHING LATCHING,
NON-
LATCHING
ALARM DELAY Number of readings above alarm
limit required to trigger alarms 5 1 - 20
BC FILT SIZE
Boxcar filter length for
concentration filter. Higher
values will smooth out the
concentration readings but will
also increase the response time.
30 1 - 30
FLOW SWITCH
ENABLE Enables/disables operation of
optional flow switch OFF ON, OFF
COMM ADDRESS MODBUS address for COM
ports. 1 1 - 247
UNIT Concentration unit-of-measure
setting. PPM PPM, WT%,
g/Nm3
SCROLL DELAY Adjustment for display scroll
delay. Higher values result in
slower scrolling speed. 20 6 - 100
AZERO ENABLE Enables/Disables auto-zero
calibration ON ON/OFF
AZERO INTERVAL Time interval for performing the
auto-zero calibration (minutes) 6 3 - 1440
GAS CARRIER
WEIGHT
Molecular weight of carrier gas
(ie. feed gas for ozone
generator.) This value is only
used for the calculation of wt%
concentration units. It is not
used for PPM or g/Nm3
calculations.
32 20.0 – 40.0
ANALOG RANGE Full-scale concentration range.
Used to scale the analog output. 5000 5 – 50 g/Nm3
1000 – 9999 PPM
0.5 – 5.0 wt%