Communications Setup and Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual
112

6.3.2. MANUALLY CONFIGURING THE NETWORK IP ADDRESSES

There are several circumstances when you may need to configure the interface

settings of the analyzer’s Ethernet card manually:

your LAN is not running a DHCP software package
the DHCP software is unable to initialize the analyzer’s interface
you wish to create a static IP (recommended)
1. Connect a cable from the analyzer’s Ethernet port to a Local Area Network
(LAN) or Internet port.
2. Access Ethernet configuration through: SETUP>MORE>COMM>INET.
3. Follow the setup sequence as shown in the illustrations that follow, and edit
the Instrument and Gateway IP addresses and Subnet Mask to the desired
settings. (From the computer, enter the same information through an
application such as HyperTerminal).
ENTR
acc ept
ne w se tti ng s
EXIT ig n or e s
ne w se tti n
g
s
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CL K MORE EXIT
SAMPLE ENTER SETUP PASS :
8
1
8
8 1 8 ENTR EXIT
SAMPLE CO2 RNG=20.00 % CO2=XXX.XX
< TST TST > CAL SETUP
SETUP X.X COMMUNICATIONS MENU
ID INET COM1 COM2 EXIT
SETUP X.X SECO N DARY SETU P M EN U
COMM VARS DIAG EXIT
SETUP X.X DHCP: ON
ON ENTR EXIT
SETUP X.X DHCP: ON
<SET SET> EDIT EXIT
SETUP X.X DHCP: OFF
OFF ENTR EXIT
(continues in next illustration)
07276B DCN6418