Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started

51

after the 60-minute warm up period is over, investigate their cause using the

troubleshooting guidelines in Section 11.

Table 3-9: Warning Messages

MESSAGE DEFINITION
ANALOG CAL WARNING The instrument's A/D circuitry or one of its analog outputs is not calibrated.
BOX TEMP WARNING The temperature inside the T803 chassis is outside the specified limits.
CANNOT DYN SPAN2 Remote span calibration failed while the dynamic span feature was set to ON
CANNOT DYN ZERO3 Remote zero calibration failed while the dynamic zero feature was set to ON
CONFIG INITIALIZED Configuration was reset to factory defaults or was erased.
CO2 ALRM1 WARNING4 CO2 concentration alarm limit 1 exceeded
CO2 ALRM2 WARNING4 CO2 concentration alarm limit 2 exceeded
CO2 CELL TEMP WARN CO2 sensor cell temperature outside of warning limits specified by
CO2_CELL_SET variable.
DATA INITIALIZED DAS data storage was erased.
O2 ALRM1 WARNING4 O2 concentration alarm limit 1 exceeded
O2 ALRM2 WARNING4 O2 concentration alarm limit 2 exceeded
O2 CELL TEMP WARN O2 sensor cell temperature outside of warning limits specified by
O2_CELL_SET variable.
REAR BOARD NOT DET CPU unable to communicate with the motherboard.
RELAY BOARD WARN CPU is unable to communicate with the relay PCA.
SAMPLE FLOW WARN The flow rate of the sample gas is outside the specified limits.
SAMPLE PRESS WARN Sample gas pressure outside of operational parameters.
SYSTEM RESET1 The analyzer was rebooted or the CPU was reset.
1 Does not clear after power up.
2 Clears the next time successful zero calibration is performed.
3 Clears the next time successful span calibration is performed.
4 Only active if the Concentration Alarm Option is installed

07276B DCN6418