Data Acquisition System (DAS & APICOM Teledyne API T803 CO2/O2 Analyzer Operation Manual
136
7.1.5.1. EDITING DAS DATA CHANNEL NAMES
To edit the name of a DAS data channel, follow the instruction shown in Section
7.1.5 then press:
Starting at the EDIT CHANNEL MENU
SETUP X.X 0) CONC: ATIMER,3,4032,RS232
NEXT INS DEL EDIT PRNT EXIT
SETUP X.X NAME: CONC
C O N C ENTR EXIT EXIT discards the new
setting
ENTR accepts the
new setting
Press each button repeatedly to
cycle through the available
character set:
0-9, A-Z, space ’ ~ ! # $ % ^ &
* ( ) - _ = +[ ] { } < >\ | ; : , . / ?
SETUP X.X NAME: CONC
SET> EDIT EXIT
7.1.5.2. EDITING DAS TRIGGERING EVENTS
Triggering events define when and how the DAS records a measurement of any
given data channel. The most commonly used triggering events are:
ATIMER: Sampling at regular intervals specified by an automatic timer. Most
trending information is usually stored at such regular intervals, which can be
instantaneous or averaged.
EXO2ZR, EXO2SP, EXO2MP, O2SLPC (exit O2 zero, exit O2 span, O2 slope
change); EXCO2Z, EXCO2S, EXCO2M, CO2SLC (exit CO2 zero, exit CO2
span, CO2 slope change); : Sampling at the end of (irregularly occurring)
calibrations or when the response slope changes. These triggering events create
07276B DCN6418