Appendix A Models T803, 803E Appendix A Menu Trees (Referen ce: 06763C DCN6418)

A-12 Error! Unknown document property name.Error! Unknown document property name. Error!

Unknown document property name.Error! Unknown document property name.

Setup Variable Numeric

Units Default

Value Value Range Description

120 SAMP_FLOW_SET cc/m
Warnings:
80–180
0–6000 Sample flow set point for flow
calculation and warning limits.
SAMP_FLOW_SLOPE — 1 0.5–1.5
Sample flow slope correction
factor (adjusted flow = measured
flow x slope).
29.92 SAMP_PRESS_SET "Hg
Warnings:
15–35
0–100 Sample pressure set point for
pressure compensation and
warning limits.
30 BOX_SET ºC
Warnings:
8–50
5–60 Box temperature warning limits.
Set point is not used.
RS232_MODE BitFlag 0 0–65535
RS-232 COM1 mode flags. Add
values to combine flags.
1 = quiet mode
2 = computer mode
4 = enable security
8 = enable hardware
handshaking
16 = enable Hessen protocol 8
32 = enable multi-drop
64 = enable modem
128 = ignore RS-232 line errors
256 = disable XON / XOFF
support
512 = disable hardware FIFOs
1024 = enable RS-48 5 mode
2048 = even parity, 7 data bits, 1
stop bit
4096 = enable command prompt
8192 = even parity, 8 data bits, 1
stop bit
16384 = enable dedicated
MODBUS ASCII protocol
32678 = enable dedicated
MODBUS RTU or TCP protocol
BAUD_RATE — 115200 300,
1200,
2400,
4800,
9600,
19200,
38400,
57600,
115200
RS-232 COM1 baud rate.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
07276B DCN6418