Troubleshooting and Service Teledyne API T803 CO2/O2 Analyzer Operation Manual
204
QUESTION ANSWER
How do I measure the sample
flow?
Sample flow is measured by attaching a calibrated rotameter, wet
test meter, or other flow-measuring device to the sample inlet port
when the instrument is operating. The sample flow should be 120
cm3/min 10%. See Section 9.4.2.
What is the averaging time for
this instrument?
The default averaging time, optimized for ambient pollution
monitoring, is 60 seconds for stable concentrations and 10 seconds
for rapidly changing concentrations; See 12.6.10 for more
information.

11.9. TECHNICAL ASSISTANCE

If this manual and its troubleshooting / repair sections do not solve your

problems, technical assistance may be obtained from:

Teledyne API, Technical Support,
9480 Carroll Park Drive
San Diego, California 92121-5201USA
Phone: 800-324-5190 (toll free in North America)
Phone: 858-657-9800 (direct)
Fax: 858-657-9816
Email: sda_techsupport@teledyne.com
Website: http://www.teledyne-api.com/

Before you contact Teledyne API’s Technical Support, fill out the problem report

form in Appendix C, which is also available online for electronic submission at

http://www.teledyne-api.com/forms/

07276B DCN6418