Models T803, 803E Appendix A Menu Trees (Referen ce: 06763C DCN6418) Appendix A

Error! Unknown document property name.Error! Unknown document property name. A-15

Setup Variable Numeric

Units Default

Value Value Range Description

CLOCK_FORMAT —
“TIME=%H:%
M:%S” Any character
in the allowed
character set.
Up to 100
characters
long.
Time-of-day clock format flags.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
“%a” = Abbreviated weekday
name.
“%b” = Abbreviated month name.
“%d” = Day of month as decimal
number (01 – 31).
“%H” = Hour in 24-hour format
(00 – 23).
“%I” = Hour in 12-hour format (01
– 12).
“%j” = Day of year as decimal
number (001 – 366).
“%m” = Month as decimal
number (01 – 12).
“%M” = Minute as decimal
number (00 – 59).
“%p” = A.M./P.M. indicator for
12-hour clock.
“%S” = Second as decimal
number (00 – 59).
“%w” = Weekday as decimal
number (0 – 6; Sunday is 0).
“%y” = Year without century, as
decimal number (00 – 99).
“%Y” = Year with century, as
decimal number.
“%%” = Percent sign.
ALARM_TRIGGER Cycles 3 1–100
Number of times concentration
must exceed limit to trigger
alarm.
07276B DCN6418