Basic Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual

62

Table 6-5: Secondary Setup Mode Features and Functions

MODE OR FEATURE MENU
ITEM DESCRIPTION MANUAL
SECTION
External Communication
Channel Configuration COM
Used to set up and operate the analyzer’s various external
I/O channels including RS-232, RS-485, modem
communication and/or Ethernet access.
5.7
System Status Variables VARS
Used to view various variables related to the instruments
current operational status
Changes made to any variable will not be recorded in
the instrument’s memory until the ENTR button is
pressed.
Pressing the EXIT button ignores the new setting.
5.5
System Diagnostic
Features
and
Analog Output
Configuration
DIAG
Used to access a variety of functions that configure, test or
diagnose problems with a variety of the analyzer’s basic
systems.
Most notably, the menus to configure the output signals
generated by the instruments Analog outputs are located
here.
5.9
Alarm Limit Configuration1 ALRM Used to turn the instrument’s two alarms on and off as well
as set the trigger limits for each. 5.11
1 Alarm warnings only present when optional concentration alarm relay package is installed.