Basic Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual

60

Table 4-2: Test Functions Defined

PARAMETER DISPLAY TITLE UNITS DEFINITION
Range
RNG
RN1
RN2

%

The full scale limit at which the reporting range of the analyzer is
currently set. THIS IS NOT the Physical Range of the instrument.
See Section 5.4.1 for more information.
Stability STB %
Standard deviation of CO2 concentration readings. Data points are
recorded every ten seconds using the last 25 data points. This
function can be reset to show O2 or CO2 stability.
Sample Pressure PRES In-Hg-A The absolute pressure of the Sample gas as measured by a
pressure sensor located inside the sample chamber.
Sample Flow SAMP FL cm3/min Sample mass flow rate as measured by the flow rate sensor in the
sample gas stream.
CO2 Sensor
Slope CO2 SLOPE - CO2 slope, computed during zero/span calibration.
CO2 Sensor
Offset CO2 OFST - CO2 offset, computed during zero/span calibration.
Box Temperature BOX TEMP C The temperature inside the analyzer chassis.
CO2 Cell
Temperature
CO2 CELL
TEMP C The current temperature of the CO2 sensor measurement cell.
O2 Cell
Temperature O2 CELL TEMP C The current temperature of the O2 sensor measurement cell.
O2 Sensor Slope O2 SLOPE - O2 slope, computed during zero/span calibration.
O2 Sensor Offset O2 OFST - O2 offset, computed during zero/span calibration.
Current Time TIME - The current time. This is used to create a time stamp on DAS
readings, and by the AUTOCAL feature to trigger calibration events.

4.3. CALIBRATION MODE

The T803 will switch into calibration mode when the user presses the CAL

button. In this mode the user can, in conjunction with introducing zero or span

gases of known concentrations into the analyzer, cause it to adjust and recalculate

the slope (gain) and offset of the its measurement range. This mode is also used

to check the current calibration status of the instrument.

Section 9 provides more information about setting up and performing standard

calibration operations or checks.

Note It is recommended that span calibration be performed at 80% of full scale of

the analyzer’s currently selected reporting range.

EXAMPLES:

If the reporting range is set for 0 to 50%, an appropriate span point would

be 40%.

If the of the reporting range is set for 0 to 100%, an appropriate span point

would be 80%.

07276B DCN6418