Teledyne API T803 CO2/O2 Analyzer Operation Manual Data Acquisition System (DAS & APICOM
135

7.1.5. SETUP DAS EDIT: ACCESSING THE DAS EDIT MODE

DAS configuration is most conveniently done through the APICOM remote

control program. The following list of touchscreen buttons shows how to edit

using the front panel.

EDIT Channel Menu
Main iDAS Menu
SETUP X.X PRIMARY SETUP MENU
CFG ACAL DAS RNGE PASS CLK MORE EXIT
SETUP X.X DATA ACQUISITION
VIEW EDIT EXIT
SETUP X.X 0) CONC: ATIMER,3,4032,RS232
NEXT INS DEL EDIT PRNT EXIT
SETUP X.X ENTER PASSWORD:818
8 1 8 ENTR EXIT
Enters EDIT mode for the selected channel
PRINT
PREV
INS
NEXT
DEL

Button

Exports the configuration of all data channels to the
RS-232 interface
Selects the previous data channel in the list
Inserts a new data channel into the list BEFORE the
selected channel
Selects the next data channel in the list
Deletes the currently selected data channel

FUNCTION

Buttons only appear when applicable

DAS EDIT – Touchscreen Button Functions

EDIT Enters EDIT mode
SAMPLE CO2 RNG=20.00 % CO2=XXX.XX
<TST TST> CAL MSG SETUP

When editing the data channels, the Param field of the display indicates some of

the DAS configuration parameters.

For example, the display line, 0) CONC: ATIMER, 3, 4032, RS232, translates

to the following configuration:

0 Channel No.
CONC Channel Name
ATIMER Trigger Event
3 Parameters – number of parameters included in this channel
4032 Event – number of data points this channel is set up to store
RS232 Port via which values automatically reported