Models T803, 803E Appendix A Menu Trees (Referen ce: 06763C DCN6418) Appendix A

Error! Unknown document property name.Error! Unknown document property name. A-13

Setup Variable Numeric

Units Default

Value Value Range Description

MODEM_INIT —
“AT Y0 &D0
&H0 &I0 S0=2
&B0 &N6 &M0
E0 Q1 &W0”
Any character
in the allowed
character set.
Up to 100
characters
long.
RS-232 COM1 modem
initialization string. Sent verbatim
plus carriage return to modem on
power up or manually. Enclose
value in double quotes (“) when
setting from the RS-232
interface.
RS232_MODE2 BitFlag 0 0–65535 RS-232 COM2 mode flags.
(Same settings as
RS232_MODE.)
BAUD_RATE2 — 19200 300,
1200,
2400,
4800,
9600,
19200,
38400,
57600,
115200
RS-232 COM2 baud rate.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
MODEM_INIT2 —
“AT Y0 &D0
&H0 &I0 S0=2
&B0 &N6 &M0
E0 Q1 &W0”
Any character
in the allowed
character set.
Up to 100
characters
long.
RS-232 COM2 modem
initialization string. Sent verbatim
plus carriage return to modem on
power up or manually. Enclose
value in double quotes (“) when
setting from the RS-232
interface.
RS232_PASS Password 940331 0–999999 RS-232 log on password.
MACHINE_ID ID 802 0–9999 Unique ID number for instrument.
COMMAND_PROMPT — “Cmd>
Any character
in the allowed
character set.
Up to 100
characters
long.
RS-232 interface command
prompt. Displayed only if enabled
with RS232_MODE variable.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
TEST_CHAN_ID — NONE

NONE,

SAMPLE

PRESSURE

,

SAMPLE

FLOW,

O2 CELL

TEMP 3,

CO2 CELL

TEMP 1,

CHASSIS

TEMP

Diagnostic analog output ID.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
07276B DCN6418