Trace Oxygen Analyzer Operation 4
4-3
Teledyne Analytical Instruments
4.3 The System Function
The subfuctions of the System function are described below. Specific
procedures for their use follow the descriptions:
Auto-Cal: Used to define an automatic calibration sequence
and/or start an Auto-Cal.
PSWD: Security can be established by choosing a 5 digit
password (PSWD) from the standard ASCII character set. (See
Installing or Changing the Password, below, for a table of
ASCII characters available.) Once a unique password is assigned
and activated, the operator MUST enter the UNIQUE password
to gain access to set-up functions which alter the instrument's
operation, such as setting the instrument span or zero setting,
adjusting the alarm setpoints, or defining analysis ranges.
Figure 4-1: Hierarchy of Functions and Subfunctions
Set Instrument
Span
Set Instrument
Zero
Set Alarm
Setpoints
Confrigure Mode
of Alarm
Operation
Define Analysi s
Ranges
Perform Ox ygen
Analysis of
the Sample
ANALYZE
SYSTEM
SPAN
ZERO
ALARMS
RANG
E
TRAK/HLD
Perform
Self-Diagnostic
Test
Initiate
Automatic
Calibration
Set Password
Logout
Show Negative