Trace Oxygen Analyzer

Appendix

 

 

 

 

 

 

 

 

 

Appendix

A-1 Specifications

Packaging:

General Purpose

 

• Flush panel mount (Standard).

 

• Relay rack mount. Contains either one or

 

two instruments in one 19" relay rack

 

mountable plate (Optional).

Sensor:

Teledyne B-2CXL trace analysis Micro-Fuel

 

Cell.

Cell Block:

316 stainless steel.

Sample System:

All wetted parts of 316 stainless steel.

90 % Response Time:

65 seconds at 25 °C (77 °F) on 10 ppm range

 

100 seconds on 1 ppm range

Ranges:

Three user definable ranges from 0–1 ppm to 0–

 

250,000 ppm, plus air calibration range of 0-

 

250,000 ppm (25 %).

 

Autoranging with range ID output.

Alarms:

One system-failure alarm contact to detect

 

power failure or sensor-zero failure.

 

Two adjustable concentration threshold alarm

 

contacts with fully programmable setpoints.

Displays:

2-line by 20-character, VFD screen, and one 5

 

digit LED display.

Digital Interface:

Full duplex RS-232 communications port.

Teledyne Analytical Instruments

A-1

Page 67
Image 67
Teledyne 3000TA-XL-EU, Trace Oxygen Analyzer operating instructions Specifications