1 Introduction

Model 3000TA-XL-EU

 

 

 

 

 

 

 

 

 

RS-232 Port

Serial digital concentration signal output

 

 

and control input.

Remote Probe

Used in the 3000TA-XL for controlling

 

 

external solenoid valves only.

Remote Span/Zero Digital inputs allow external control of analyzer calibration.

Calibration Contact To notify external equipment that instrument is being calibrated and readings are not monitoring sample.

Range ID Contacts Four separate, dedicated, range relay contacts. Low, Medium, High, Cal.

Network I/O

Serial digital communications for local

 

network access. For future expansion.

 

Not implemented at this printing.

Note: If you require highly accurate Auto-Cal timing, use external Auto-Cal control where possible. The internal clock in the Model 3000TA-XL is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day.