Trace Oxygen Analyzer

Maintenance 5

 

 

 

 

 

 

 

 

 

1.Remove power to the instrument by unplugging the power cord at the power source.

2.Open the front panel door by pressing the release button on the top right corner of the door all the way in with a narrow gauge tool.

3.With one hand placed underneath the cell block ready to catch the Micro-Fuel cell , lift up on the stainless steel gate in front of the cell block. This releases the cell and cell holder from the block. The cell and holder will fall out in your hand.

Needle Valve

Lift Up

Cell Block

Gate

Micro-Fuel Cell

Cell Adaptor

Figure 5-1: Removing the Micro-Fuel

O-Ring

Cell Holder