Trace Oxygen Analyzer Maintenance 5
5-3
Teledyne Analytical Instruments
1. Remove power to the instrument by unplugging the power cord
at the power source.
2. Open the front panel door by pressing the release button on the
top right corner of the door all the way in with a narrow gauge
tool.
3. With one hand placed underneath the cell block ready to catch
the Micro-Fuel cell , lift up on the stainless steel gate in front of
the cell block. This releases the cell and cell holder from the
block. The cell and holder will fall out in your hand.
Figure 5-1: Removing the Micro-Fuel
Cell Blo ck
Micro-Fuel Cell
O-Ring
Cell Holder
Lift Up
Gate
Cell Adaptor
Needle Valve