Trace Oxygen Analyzer | Operation 4 | |
|
|
|
|
|
|
|
|
|
In the first line, TRACK or HOLD should be blinking. The operator can toggle between TRACK and HOLD with the Up or Down keys. When TRACK is selected, the analog outputs
When HOLD is selected, the analog outputs
The concentration alarms freeze on their last state before entering
calibration regardless of selecting HOLD or TRACK. But, when HOLD is selected the concentration alarms will remain frozen for the time displayed in the second line of the TRAK/HLD menu after the analyzer returns to the Analyze mode.
The factory default is three minutes, but the delay time is program- mable. To adjust to delay time use the Left or Right arrow keys. When the time displayed on the second line blinks, it can be adjusted by Pressing the Up or Down keys to increase or decrease its value. The minimum delay is 1 minute, the maximum is 30.
This preference is stored in
4.3.2 Setting up an Auto-Cal
When proper automatic valving is connected (see chapter 3, installa- tion), the Analyzer can cycle itself through a sequence of steps that automati- cally calibrates the instrument.
Note: If you require highly accurate
To setup an
CAUTION:
We do not recommend frequent Zero adjustments of the cell. A newly installed cell may take