Trace Oxygen Analyzer

Maintenance 5

 

 

 

 

 

 

 

 

 

Maintenance

5.1Routine Maintenance

Aside from normal cleaning and checking for leaks at the gas connec- tions, routine maintenance is limited to replacing Micro-Fuel cells and fuses, and recalibration. For recalibration, see Section 4.4 Calibration.

WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS

MANUAL.

5.2Cell Replacement

The B-2CXL Micro-Fuel Cell is a sealed electrochemical transducer with no electrolyte to change or electrodes to clean. When the cell reaches the end of its useful life, it is replaced. The spent fuel cell should be dis- carded according to local regulations. This section describes fuel cell care as well as when and how to replace it.

5.2.1 Storing and Handling Replacement Cells

To have a replacement cell available when it is needed, TAI recom- mends that one spare cell be purchased 9-10 months after commissioning the 3000TA-XL, or shortly before the end of the cell's one year warranty period.

CAUTION: Do not stockpile cells. The warranty period starts on the day of shipment.

The spare cell should be carefully stored in an area that is not subject to large variations in ambient temperature (75 °F nominal) or to rough handling.

Teledyne Analytical Instruments

5-1

Page 59
Image 59
Teledyne 3000TA-XL-EU operating instructions Routine Maintenance, Cell Replacement, Trace Oxygen Analyzer Maintenance