Trace Oxygen Analyzer

Operation 4

 

 

 

 

 

 

 

 

 

Zero:Settling:Man

<ENT> To Begin

Press Enter to begin the zero calibration. After a few seconds the first of five zeroing screens appears. The number in the upper left hand corner is the first-stage zero offset. The microprocessor samples the output at a prede- termined rate. It calculates the differences between successive samplings and displays the rate of change as Slope= a value in parts per million per second (ppm/s).

####ppm Zero

Slope=#### ppm/s

NOTE: It takes several seconds for the true Slope value to display. Wait about 10 seconds. Then, wait until Slope is sufficiently close to zero before pressing Enter to finish zeroing .

Generally, you have a good zero when Slope is less than 0.05 ppm/s for about 30 seconds. When Slope is close enough to zero, press Enter. In a few seconds, the screen will update.

Once zero settling is completed, the information is stored in the microprocessor, and the instrument automatically returns to the Analyze mode.

4.4.1.3Cell Failure

Cell failure in the 3000TA-XL is usually associated with inability to zero the instrument down to a satisfactorily low ppm reading. When this occurs, the instrument will still Zero, but the 3000TA-XL system alarm trips, and the LCD displays a failure message.

#.#ppm Anlz

CELL FAIL/ ZERO HIGH

Before replacing the cell:

a.Check your span gas to make sure it is within specifications.

b.Check for leaks up-stream from the cell, where oxygen may be leaking into the system.

If there are no leaks and the span gas is within specification, replace the cell as described in chapter 5, Maintenance.

Teledyne Analytical Instruments

4-13

Page 49
Image 49
Teledyne 3000TA-XL-EU, Trace Oxygen Analyzer operating instructions Cell Failure