3 Installation Model 3000TA-XL-EU
3-2 Teledyne Analytical Instruments
Figure 3-1: Front Panel of the Model 3000TA-XL
All operator controls, except the flow control needle valve, are mounted
on the control panel, which is hinged on the left edge and doubles as the
door that provides access to the sensor and cell block inside the instrument.
The door latch is spring loaded and will swing open when the button in the
center of the latch (upper right corner) is pressed all the way in with a narrow
gauge tool (less than 0.18 inch wide). Allow clearance for the door to open
in a 90-degree arc of radius 7.125 inches. See Figure 3-2.
Figure 3-2: Required Front Door Clearance
3.3 Rear Panel Connections
Figure 3-3 shows the Model 3000TA-XL rear panel. There are ports
for gas inlet and outlet, power, communication, and both digital and analog
concentration output.
10"
6.7"