3 Installation

Model 3000TA-XL-EU

 

 

 

 

 

 

 

 

 

Can be configured as failsafe or nonfailsafe.

Can be configured as latching or nonlatching.

Can be configured out (defeated).

Threshold Alarm 2:

• Can be configured as high (actuates when concen-

 

tration is above threshold), or low (actuates when

 

concentration is below threshold).

 

• Can be configured as failsafe or nonfailsafe.

 

• Can be configured as latching or nonlatching.

 

• Can be configured out (defeated).

System Alarm:

Actuates when DC power supplied to circuits is

 

unacceptable in one or more parameters. Permanently

 

configured as failsafe and latching. Cannot be de-

 

feated. Actuates if self test fails.

 

(Reset by pressing

button to remove power. Then

 

press

again and any other button EXCEPT

System to resume.

Further detail can be found in chapter 4, section 4-5.

Table 3-2: Alarm Relay Contact Pins

Pin Contact

45Threshold Alarm 1, normally closed contact

28Threshold Alarm 1, moving contact

46Threshold Alarm 1, normally open contact

42Threshold Alarm 2, normally closed contact

44Threshold Alarm 2, moving contact

43Threshold Alarm 2, normally open contact

36System Alarm, normally closed contact

20System Alarm, moving contact

37System Alarm, normally open contact

Digital Remote Cal Inputs: Accept 0 V (off) or 24 V dc (on) inputs for remote control of calibration. (See Remote Calibration Protocol below.) See Table 3-3 for pin connections.

Zero:

Floating input. 5 to 24 V input across the + and – pins puts

 

the analyzer into the Zero mode. Either side may be

 

grounded at the source of the signal. 0 to 1 volt across the

 

terminals allows Zero mode to terminate when done. A

 

synchronous signal must open and close the external zero

 

valve appropriately. See Remote Probe Connector. (The –C

 

option internal valves operate automatically.)

 

 

3-6

Teledyne Analytical Instruments

Page 30
Image 30
Teledyne Trace Oxygen Analyzer, 3000TA-XL-EU operating instructions Pin Contact