A-12
Appendix Model 3000TA-XL-EU
Teledyne Analytical Instruments
Micro fuel cell
Protective Film/
Seal Cover
(Located this side)
Cell Block
Lift up Gate to
removed Cell
Holder.
Cell Holder
O-Ring
It is important to minimize the amount
of time a Teledyne Trace Oxygen
sensor is exposed to air during the
installation process. The quicker the
sensor can be installed into the unit,
the faster your sensor will recover to
low oxygen measurement levels.
CAUTION: Do not touch the sensing
surface of the cell. It is
covered with a delicate
Teflon membrane that can
leak when punctured. The
sensor must be replaced if
the membrane is damaged.
Check O-ring in the base of the cell holder
before installing a new cell. Replace it if
worn or damaged.
A7 - Installling and Replacing a new Insta-TraceMicro-fuel Cell

Read completely before beginning the procedure