VXI Technology, Inc.
Bit 14 - Measures 128 Ω on the 10 kΩ scale
Bit 15 - Measures 81.92 kΩ on the 100 kΩ scale
Bit 16 - Measures 81.92 kΩ on the 1 MΩ scale
Bits 17 – 31 are not used and read as “0”.
For Channels
Bit 0 - Measures +0.117 V on the 0.5 V scale
Bit 1 - Measures
Bit 2 - Measures +0.945 V on the 1 V scale
Bit 3 - Measures
Bit 4 - Measures +0.945 V on the 2 V scale
Bit 5 - Measures
Bit 6 - Measures +0.945 V on the 5 V scale
Bit 7 - Measures
Bit 8 - Measures +9.45 V on the 10 V scale
Bit 9 - Measures
Bit 10 - Measures +9.45 V on the 20 V scale
Bit 11 - Measures
Bit 12 – 31 are not used and read as “0”.
Example 1
Reading 0x0000000C after a Self Test command indicated a problem during Self Test 0 and Self Test 1, measuring ±0.945 V on the 2 V scale, but no problem on all other tests. This would indicate that the front end programmable gain amplifier has a problem on the x5 (gain = 5) setting.
Example 2
Reading 0x00000000 after a Self Test operation would indicate that all Self Tests passed successfully.
Preset Setting Measurement Commands
The following measurement commands are executed by the microprocessor using
0x0005 = 100 Ω Range Query
0x000A =
0x000D = 10 kΩ range query
0x0010 =
0x0012 = 1 V Measurement
0x0013 = 2 V Measurement
0x0014 = 5 V Measurement
0x0015 = 10 V Measurement
44 | SVM2608 Programming |