www.vxitech.com

 

 

 

 

Control Register (0x08, 0x30, 0x58, 0x80, 0xA8, 0xD0) — Read & Write

 

 

 

 

 

 

 

 

Linear or FIFO Memory Mode - Determines whether the data is

 

 

 

 

 

 

 

captured in Linear (burst) or FIFO (real time) acquisition mode.

 

D10

 

LINEAR/FIFO

 

0 = Linear mode

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1 = FIFO mode

 

 

 

 

 

 

 

Pon state = 0

 

 

 

 

 

 

 

Function Setting - The digitizer is capable of measuring voltages,

 

 

 

 

 

 

 

 

resistances in 2-wire mode or resistances in 4-wire mode. Before taking a

 

 

 

 

 

 

 

 

measurement, allow for at least 5 ms for internal circuits to settle after

 

 

 

 

 

 

 

 

making changes.

 

 

D9 - D8

 

 

2WIREOHMS/

 

 

00 = Voltage mode

 

 

 

 

4WIREOHMS

 

 

01 = 4-wire resistance mode

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

10 = 2-wire resistance mode

 

 

 

 

 

 

 

 

11 = Invalid state

 

 

 

 

 

 

 

 

Note: These bits are only utilized by Channels 0 – 3. These bits are

 

 

 

 

 

 

 

 

unused by Channels 4 and 5.

 

 

 

 

 

 

 

 

Attenuation and Gain Setting - Valid attenuation settings are x1 and

 

 

 

 

 

 

 

x10. Valid gain settings are 1, 2, 5, and 10. Note that there is only one

 

 

 

 

 

 

 

current source shared among all four channels, therefore, only one

 

 

 

 

 

 

 

resistance measurement can be made at any one time. Before taking a

 

 

 

 

 

 

 

measurement, allow for at least 5 ms for internal circuits to settle after

 

 

 

 

 

 

 

making changes.

 

 

 

 

 

 

 

VOLTAGE MODE (for channels 0-3)

 

 

 

 

 

 

011

= 1.0 V

 

 

 

 

 

 

010

= 2.0 V

 

 

 

 

 

 

001

= 5.0 V

 

 

 

 

 

 

000

= 10.0 V

 

 

 

 

 

 

110

= 20.0 V

 

 

 

 

 

 

101

= 50.0 V

 

 

 

 

 

 

100

= Invalid state

 

D7 – D5

 

ATTN-GAIN1-GAIN0

111

= Invalid state

 

 

 

 

 

 

VOLTAGE MODE (for channels 4-5) 011 = 0.5 V

001 = 1.0 V

000 = 2.0 V

111 = 5.0 V

101 = 10.0 V

100 = 20.0 V

010 = Invalid state

110 = Invalid state

RESISTANCE MODE (Valid for channels 0-3) 011 = 100 Ω; 2 mA current source setting 000 = 1 kΩ; 2 mA current source setting

000 = 10 kΩ; 200 μA current source setting 000 = 100 kΩ; 20 μA current source setting 000 = 1 MΩ; 2 μA current source setting