VXI Technology, Inc.
The aforementioned errors are reported for the following reasons:
NO_ERROR | There are no errors in the queue. |
ERROR_UNKNOWN_COMMAND | An unknown command was sent to the |
| microprocessor. |
ERROR_PRE_GT_SIZE | The value programmed in the |
| Points register is greater than the value |
| programmed in the Sample Points |
| register. |
ERROR_RESISTANCE_OVER_RANGE | The user attempted to measure a |
| resistance greater than the range. |
ERROR_UNSTABLE_RESISTANCE | In |
| resistance is not stable. |
ERROR_UNSTABLE_VOLTAGE | In |
| is not stable. |
ERROR_MULTIPLE_TEST_SOURCES | The user attempted to connect several |
| sources on the Test Bus. |
ERROR_NONVOL_READ | There was an error when trying to read |
| the |
ERROR_NONVOL_WRITE | There was an error when trying to write |
| to the |
ERROR_NONVOL_DEFAULTED | There was an error when trying to read |
| the default values from the |
| calibration memory. |
ERROR_FLASH_BURN | Error trying to burn data in the FLASH (a |
| common error here would occur when |
| loading the firmware for U1 and issuing |
| the write flash commands for U20, or |
| vice versa). |
ERROR_INTERNAL_SOFTWARE | The microprocessor encountered an |
| internal software error. |
ERROR_QUEUE_OVFL | There were more then five errors in the |
| ERROR queue. |
48 | SVM2608 Programming |