VXI Technology, Inc.

20 SVM2608 Introduction

SVM2608 SPECIFICATIONS

GENERAL SPECIFICATIONS

NUMBER OF CHANNELS

4

SAMPLING RATE

Range
Resolution
Accuracy
0.59 Samples/s to 100.0 kSamples/s
100 ns
1%

INPUT POWER

+5 V dc
+12 V dc
-12 V dc
500 mA
300 mA
300 mA

VMEBUS INTERFACE

Address mode
Data transfer mode
A32
D16 or D32

MEMORY

1 MSamples per channel

VOLTAGE MEASUREMENT

Range
Resolution
Accuracy
±1.0 V, ±2.0 V, ±5.0 V, ±10.0 V, ±20.0 V, ±50.0 V
1/215 of full scale
1%

INPUT IMPEDANCE

20 & 50 V Range
1, 2, 5 & 10 V Range
200 k
> 10 M

RESISTANCE MEASUREMENT*

Range
Resolution
Accuracy
100 , 1 k, 10 k, 100 k, 1 M, 100% over range*
1/6,553.6 of scale
1%

*Note: Resistance measurements can only be made one channel at a time.

All resistance measurements can be made accurately up to +199% of the set range.

INPUT FILTER

20 kHz (-3 dB)

TRIGGER LEVELS

Internal
Range
Trigger level resolution
Level accuracy
External
Range
Trigger level resolution
Level accuracy
trigger level determined by selected voltage range (see Voltage Measurement above)
Range/211
1%
-10 V to +10 V
4.88 mV
1%

DELAYED TRIGGER

Range
Resolution
0 samples to (232 -1) samples
1 sample

TIMEOUT

Range
Resolution
10 µs to 227 hrs
10 µs to 100 s

WARM-UP TIME

10 min

MTBF

80,000 hrs (based on 20% Ground Mobile, 80% Ground Fixed environment, T 52°C)