VXI Technology, Inc.

 

 

SVM2608 SPECIFICATIONS

 

 

 

GENERAL SPECIFICATIONS

 

 

NUMBER OF CHANNELS

 

 

 

4

SAMPLING RATE

 

 

Range

 

0.59 Samples/s to 100.0 kSamples/s

Resolution

 

100 ns

Accuracy

1%

INPUT POWER

 

 

+5 V dc

 

500 mA

+12 V dc

 

300 mA

-12 V dc

 

300 mA

VMEBUS INTERFACE

 

 

Address mode

 

A32

Data transfer mode

 

D16 or D32

MEMORY

 

 

 

 

1 MSamples per channel

VOLTAGE MEASUREMENT

 

 

Range

 

±1.0 V, ±2.0 V, ±5.0 V, ±10.0 V, ±20.0 V, ±50.0 V

Resolution

 

1/215 of full scale

Accuracy

1%

INPUT IMPEDANCE

 

 

20 & 50 V Range

 

200 kΩ

1, 2, 5 & 10 V Range

 

> 10 MΩ

RESISTANCE MEASUREMENT*

 

 

Range

 

100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 100% over range*

Resolution

1/6,553.6 of scale

Accuracy

1%

*Note: Resistance measurements can only be made one channel at a time.

All resistance measurements can be made accurately up to +199% of the set range.

INPUT FILTER

 

 

 

 

20 kHz (-3 dB)

TRIGGER LEVELS

 

 

Internal

 

trigger level determined by selected voltage range (see Voltage Measurement above)

Range

 

Trigger level resolution

 

Range/211

Level accuracy

1%

External

 

 

Range

 

-10 V to +10 V

Trigger level resolution

 

4.88 mV

Level accuracy

1%

DELAYED TRIGGER

 

 

Range

 

0 samples to (232 -1) samples

Resolution

 

1 sample

TIMEOUT

 

 

Range

 

10 µs to 227 hrs

Resolution

 

10 µs to 100 s

WARM-UP TIME

10 min

MTBF

80,000 hrs (based on 20% Ground Mobile, 80% Ground Fixed environment, T ≥ 52°C)