Teledyne Model 300EU ANALOG OUTPUT Connections, M300EU, Analog Output Data Type Default Settings

Page 15
3.3.2. ANALOG OUTPUT Connections

M300EU

GETTING STARTED

(Addendum to M300EU Manual - P/N 04145)

 

3.3.2. ANALOG OUTPUT Connections

The analog outputs for the M300EU are different from those of the M300E/EM. Unlike the M300E/EM which can include options for measuring both O2 and CO, the M300EU is a single gas analyzer (CO) and therefore does not require the fully configurable version of the analog outputs. The following information replaces that found in Section 3.1.2.2 of the M300E/EM Operators Manual (P/N 04288).

The M300EU is equipped with several analog output channels accessible through a connector on the back panel of the instrument.

Channels A1 and A2 output a signal that is proportional to the CO concentration of the sample gas.

The default analog output voltage setting of these channels is 0 to 5 VDC with a reporting range of 0 to 500 ppb.

An optional Current Loop output is available for each.

The output labeled A4 is special. It can be set by the user to output any one a variety of diagnostic test functions.

The default analog output voltage setting of these channels is also 0 to 5 VDC.

See Section 6.2.2 for a list of available functions.

There is NO optional Current Loop output is available for Channel A4.

The following table lists the default settings for each of these channels.

Table 3-1:

Analog Output Data Type Default Settings

 

 

 

 

 

 

 

 

PARAMETER

 

 

CHANNEL DEFAULT SETTING

 

 

A1

 

A2

A3

 

A41

DATA TYPE1

CONC1

 

CONC2

Not

 

TEST

 

 

 

 

Available

 

CHANNEL

RANGE

 

0 – 5

VDC2

 

 

 

OVERRANGE

 

ON

 

 

 

REC OFS

 

0 mVDC

 

 

 

AUTO CAL.

 

ON

 

 

 

CALIBRATED

 

NO

 

 

 

OUTPUT

 

ON

 

 

 

1 See Table A-6 of M300E/EM Appendix A for definitions of these iDAS data types 2 Optional current loop outputs are available for analog output channels A1& A2.

To access these signals attach a strip chart recorder and/or data-logger to the appropriate analog output connections on the rear panel of the analyzer. Pin-outs for the analog output connector are:

ANALOG OUT

A1 A2 A3 A4

+- + - + - + -

Figure 3-3: M300EU Analog Output Connector

05514 Rev A1

7

Image 15
Contents 9480 CARROLL PARK DRIVE SAN DIEGO, CALIFORNIA USA ADDENDUM TO M300E/EM MANUAL P/NMODEL 300EU CO ANALYZER WITH AUTO-REFERENCE TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION DIVISION TAPIPage GENERAL SAFETY HAZARD SAFETY MESSAGESUSER NOTES M300EUAddendum to M300E/EM Manual P/N TABLE OF CONTENTS LIST OF TABLES LIST OF FIGURESAPPENDIX A-4 Model 300EU Signal I/O Definitions, Revision K.6 LIST OF APPENDICESAddendum to M300E/EM Manual P/N any time the instrument is in SETUP mode 1. INTRODUCTION1.1. REFERENCE NUMBERING CONVENTION 1.2. M300EU OVERVIEWAddendum to M300E/EM Manual P/N 05514 Rev A1 2. SPECIFICATIONS, APPROVALS AND WARRANTY2.1. SPECIFICATIONS M300EU Basic Unit Specifications05831 Rev 2.4. WARRANTY2.2. EPA EQUIVALENCY DESIGNATION 2.3. CE MARK COMPLIANCEOPTICAL BENCH 3. GETTING STARTED3.1. UNPACKING THE M300EU 3.2. M300EU ANALYZER LAYOUTFigure 3-2 M300EU Internal Gas Flow 3.3.1. POWER CONNECTION3.3. ELECTRICAL CONNECTIONS M300EU3.3.2. ANALOG OUTPUT Connections Analog Output Data Type Default SettingsFigure 3-3 M300EU Analog Output Connector Output # 3.3.3. CONNECTING THE STATUS OUTPUTSAnalog Output Pin Outs Table 3-3 Status Output Pin AssignmentsMESSAGE 3.4. INITIAL OPERATION OF THE M300EU3.4.1. STARTUP AND WARM UP OF THE M300EU 3.4.2. WARNING MESSAGESThe Test functions available from the front panel of the M300EU are 3.4.3. FUNCTIONAL CHECK3.5.2.1. Calibration Gases 3.5.2. BASIC CO CALIBRATION SETUP3.5. INITIAL CALIBRATION 3.5.1. PRE-CALIBRATION STABILIZATIONSTEP 3 - SET CO SPAN GAS CONCENTRATION 3.5.3. BASIC CO CALIBRATION PROCEDURESTEP 1 - SET LIMITS STEP 2 - DILUTION RATIO4. FREQENTLY ASKED QUESTIONS & GLOSSARY Addendum to M300E/EM Manual P/N 300E 5. OPTIONAL HARDWARE AND SOFTWAREOPTION DESCRIPTIONM300EU Internal Pneumatic Flow- Zero/Span Valves OPT 50A 5.1. CALIBRATION VALVE OPTIONS FOR THE M300EU5.1.1. ZERO SPAN VALVES OPT 50A TOLL-FREE FAX TEL E-MAIL api-sales@teledyne.com5.1.2. ZERO SPAN VALVES WITH IZS OPT 51B SAMPLE CHAMBERScrubber 5.1.3. Zero Span Valves with IZS and Shutoff VALVE OPT 51B 6. M300EU OPERATING INSTRUCTIONS reference gas path 6.2. OPERATING MODESM300EU Operating Modes sample gas stream is being routed through the CO scrubber of the autoAUTO-REF RANGE = 50.000 PPM 6.2.1. AUTO-REFERENCE MODE AREFSETUP Mode FieldParameter 6.2.2. TEST FUNCTIONSUnits Table 6-2 Test Functions DefinedANALOG OUT 6.3. SETUP Æ RNGE ANALOG OUTPUT REPORTING RANGE CONFIGURATION6.3.1. PHYSICAL RANGE VERSUS ANALOG OUTPUT REPORTING RANGES 6.3.2. ANALOG OUTPUT RANGES FOR CO CONCENTRATIONAUTO DEFAULT SETTINGSSNGL DUALThis is the default reporting range mode for the analyzer To set the ranges press following keystroke sequence M300EU OPERATING INSTRUCTIONS Consult your local regulations for the STP used by your agency 6.3.6. SETUP Æ RNGE Æ UNIT SETTING THE REPORTING RANGE UNIT TYPEweight/volume units CO ppb x 1.25 = CO ug/m3 CO ppm x 1.25= CO mg/m3 Pressure STP6.4.1. ADJUSTING THE A-REF MODE CYCLE TIME To manually start an auto-reference measurement cycle, press 6.4.2. FORCING AN AUTO-REFERENCE MEASUREMENT05514 Rev A1 DESCRIPTION 6.5. SETUP Æ MORE Æ VARS INTERNAL VARIABLES VARSTable 6-3 Test Functions Defined VARIABLEDIAG AIO 6.6. SETUP Æ MORE Æ DIAG USING THE DIAGNOSTICS FUNCTIONSDIAG I/O DIAG AOUTDIAG - Analog I/O Functions 6.7. ANALOG OUTPUT CONFIGURATION6.7.1. ANALOG I/O CONFIGURATION THE ANALOG I/O CONFIGURATION SUBMENU4. Choose a Test Channel parameter to be output on the channel 6.13.5.8 SETTING DATA UPDATE RATE FOR AN ANALOG OUTPUT ANALOG OUTPUT SIGNAL TYPE AND RANGE SELECTIONTURNING THE ANALOG OUTPUT OVER-RANGE FEATURE ON/OFF 6.13.5.7 SETTING THE REPORTING RANGE SCALE FOR AN ANALOG OUTPUTAUTOMATIC ANALOG OUTPUT CALIBRATION 6.7.2. CALIBRATION OF THE ANALOG OUTPUTSMANUAL CALIBRATION OF ANALOG OUTPUT CONFIGURED FOR VOLTAGE RANGES ENABLING OR DISABLING THE AUTO-CAL FOR AN INDIVIDUAL ANALOG OUTPUTZERO 6.7.3. SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4Test Channels Functions Available on the M300EU’s Analog Output TEST CHANNELTable 6-7 M300EU Hessen GAS ID List 6.7.4. HESSEN PROTOCOLDefault Hessen Status Bit Assignments STEP3 - ZERO/SPAN CALIBRATION 7. CALIBRATION PROCEDURESmanual for troubleshooting tips MANUAL CALIBRATION REQUIRED EQUIPMENT, SUPPLIES AND EXPENDABLESMANUAL CALIBRATION CHECKS WITH ZERO/SPAN VALVES 7.9.1 SECTION O2 SENSOR CALIBRATION PROCEDURE7.9.2 CO2 SENSOR CALIBRATION PROCEDURE MANUAL CALIBRATION WITH ZERO/SPAN VALVES8. EPA PROTOCOL CALIBRATION Addendum to M300E/EM Manual P/N 9.1. PERFOMING AN AUTO-REFERENCE MEASUREMENT AFTER MAINTENANCE 9. INSTRUMENT MAINTENANCEAddendum to M300E/EM Manual P/N MR RATIO Table 9-1 M300EU Test Function RecordSTABILITY CO MEASAddendum to M300E/EM Manual P/N Figure 10-1 M300EU Gas flow during Auto-Reference Measurements 10. THEORY OF OPERATION10.1. MEASUREMENT METHOD 10.1.1. AUTO-REFERENCE RATIOThis includes calibration gases such as zero air and span gas 10.2. PNEUMATIC OPERATION10.2.1. THE A-REF CARBON MONOXIDE SCRUBBER Figure 10-2 Auto-Reference Measurement CycleFigure 10-3 Semi-Permeable Membrane Drying Process 10.2.2. THE A-REF VALVE ASSEMBLY10.2.3. THE NAFION→ DRYER Auto-Reference Valve Operating StatesThe dryer consists of a bundle of 0.030 O.D. Nafion→ tubes housed within a single large tube shell. Sample gas flows within the Nafion tubes while water vapor absorbs into the tubular membrane walls and is removed. Moisture permeating the tubing is carried away by a dry purge gas within the shell that flows countercurrent to the sample gas Aout 10.3. ELECTRONIC OPERATION10.3.1. OVERVIEW Figure 10-4 M300EU Electronic Overview Block Diagram10.3.2. THE RELAY PCA 10.3.2.1. Temperature Control of the Convection OvenFigure 10-5 M300EU Heating Control Block Diagram Figure 10-6 M300EU Oven Heater Configuration Jumpers 10.3.2.2. Oven Heaters AC Power Configuration115 VAC Configuration 230 VAC Configuration10.3.3.2. Sensor Inputs 10.3.3. MOTHERBOARD10.3.2.3. Status LED’s 10.3.3.1. A to D Conversion10.3.4. POWER DISTRIBUTION RELAY PCAFigure 10-8 M300EU Distribution Block Diagram Page PROCEDURES 11. TROUBLESHOOTING & REPAIR11.1. GENERAL NOTES MAINTENANCE PERSONNEL ONLY11.2. FAULT DIAGNOSIS WITH WARNING MESSAGES Warning Messages - Indicated FailuresTable continued on next page Sample flow rate is Warning Messages - Indicated Failures contINDICATED FAILURES 11.2.1. FAULT DIAGNOSIS WITH TEST FUNCTIONSTable 11-2 Test Functions - Indicated Failures table continuedValues outside range indicate Test Functions - Indicated FailuresVoltage displayed should change. If not 11.2.2. RELAY BOARD STATUS LED’SFaulty Connectors/Wiring Table 11-3 Relay Board Status LED Failure Indications11.3. GAS FLOW PROBLEMS 11.4. OTHER PERFORMANCE PROBLEMS11.4.1. Unexplained DRIFT 11.5.2.1. A/D Functions 11.5. SUBSYSTEM CHECHOUT11.5.1. RELAY BOARD 11.5.2. MOTHERBOARD11.6. TECHNICAL ASSISTANCE Addendum to M300E/EM Manual P/N M300EU Addendum to M300E/EM Manual APPENDIX A - Version Specific Software DocumentationAPPENDIX A-2 M300EU Setup Variables Available Via Serial I/O APPENDIX A-4 M300EU Signal I/O DefinitionsCLR1 TEST1MSG1 Figure A-2 Primary Setup Menu Except iDAS Figure A-3 Primary Setup Menu Æ iDAS Submenu Figure A-4 Secondary Setup Menu Æ COMM and VARS Submenus Figure A-5 Secondary Setup Menu Æ Hessen Protocol Submenu Press ENTR ENTER PASSWORDAuto Cal feature is OFF or the range is set for Figure A-6 DIAG MenuAPPENDIX A-2 Setup Variables For Serial I/O, Revision K.6 APPENDIX A-2 Setup Variables For Serial I/O, Revision K.6Table continues CO2FILTPCT A-10 O2FILTPCTSTDO2CELLTEMP A-11 APPENDIX A-3 Warnings and Test Functions, Revision K.6NAME MESSAGE TEXTThese gases not applicable to the M300EU Table A-5A-12 These gases not applicable to the M300EUNumber Signal NameBit or Channel DescriptionAPPENDIX A-5 M300EU iDAS Parameters Revision K.6 OVTMPW Internal box temperature #2/oven warningA-14 O2OFST A-15CO2OFS O2SLPEA-16 APPENDIX A-6 Terminal Command Designators, Revision K.605925 Rev A APPENDIX B - M300EU SPARE PARTS & EXPENDABLESAPPENDIX B - M300EU SPARE PARTS & EXPENDABLES European standard ENM300EU Addendum to M300E/EM Manual