M300EU OPERATING INSTRUCTIONS | M300EU |
| Addendum to M300E/EM Manual P/N 04288 |
6.2.2. TEST FUNCTIONS
The information found in Section 6.2.1 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s):
•The following table supersedes Table
Table 6-2: Test Functions Defined
Parameter | Display Title | Units | Meaning | |
|
|
| The | |
RANGE |
|
| Outputs is currently set. | |
- - | RANGE | PPB, PPM | • THIS IS NOT the Physical Range of the instrument. See Section 6.3.2 | |
RANGE1 | UGM, MGM | for more information. | ||
| ||||
RANGE2 |
|
| If DUAL or AUTO Range modes have been selected, two RANGE functions | |
|
|
| will appear, one for each range. | |
|
| PPB, PPM | Standard deviation of CO concentration readings. Data points are recorded | |
Stability | STABIL | every ten seconds using the last 25 data points. This function can be reset | ||
UGM, MGM | to show O2 or CO2 stability in instruments with those sensor options | |||
|
| |||
|
|
| installed. | |
|
|
|
| |
CO Measure | MEAS | MV | The demodulated, peak IR detector output during the measure portion of the | |
CFG Wheel cycle. | ||||
|
|
| ||
CO Reference | REF | MV | The demodulated, peak IR detector output during the reference portion of | |
the CFG wheel cycle. | ||||
|
|
| ||
|
|
| The result of CO MEAS divided by CO REF based on readings taken during | |
Measurement / | MR Ratio | - | the normal sample measurement portion of the | |
Reference Ratio | This ratio is the primary value used to compute CO concentration. The | |||
|
| |||
|
|
| value displayed is not linearized. | |
|
|
|
| |
|
|
| The result of CO MEAS divided by CO REF based on readings | |
AZERO RATIO | - | taken during the | ||
This ratio is the used to compute a reference correction factor | ||||
Ratio | ||||
|
| for computing the CO concentration. The value displayed is | ||
|
|
| ||
|
|
| not linearized. | |
Sample Pressure | PRES | The absolute pressure of the Sample gas as measured by a pressure | ||
sensor located inside the sample chamber. | ||||
|
|
| ||
Sample Flow | SAMPLE FL | cm3/min | Sample mass flow rate as measured by the flow rate sensor in the sample | |
|
|
| gas stream, | |
Sample | SAMP TEMP | °C | The temperature of the gas inside the sample chamber. | |
Temperature | ||||
|
|
| ||
Bench | BENCH TEMP | °C | Optical bench temperature. | |
Temperature | ||||
|
|
| ||
Wheel | WHEEL TEMP | °C | GFC wheel temperature. | |
Temperature | ||||
|
|
| ||
Box Temperature | BOX TEMP | °C | The temperature inside the analyzer chassis. | |
Oven | OVEN TEMP2 | °C | The current temperature of the circulating air inside the M300EU’s | |
Temperature |
|
| convection oven area. | |
|
| The drive voltage being supplied to the thermoelectric coolers of the IR | ||
Temp. Control | PHT DRIVE | mV | ||
Voltage |
|
| ||
|
|
| ||
|
|
|
| |
|
|
| The sensitivity of the instrument as calculated during the last calibration | |
Slope | SLOPE | - | activity. The SLOPE parameter is used to set the span calibration point of | |
|
|
| the analyzer. | |
|
|
| The overall offset of the instrument as calculated during the last calibration | |
Offset | OFFSET | - | activity. The OFFSET parameter is used to set the zero point of the | |
|
|
| analyzer response. | |
Test Channel | TEST | mV | The raw voltage being output on the analyzer’s A4 analog output. Only | |
Output | appears when the test channel is assigned a function. | |||
|
| |||
Current Time | TIME | - | The current time. This is used to create a time stamp on iDAS readings, and | |
by the AUTOCAL feature to trigger calibration events. | ||||
|
|
|
22 | 05831 Rev. 01 |