| M300EU |
|
| TROUBLESHOOTING & REPAIR |
|
| (Addendum to M300EU Manual - P/N 04145) |
|
| ||
|
| Table | Warning Messages – Indicated Failures (cont.) |
| |
|
|
|
|
|
|
| WARNING | FAULT CONDITION | POSSIBLE CAUSES |
| |
| MESSAGE |
| |||
|
|
|
|
| |
|
| Sample flow rate is | Failed sample pump |
| |
|
| Blocked sample inlet/gas line |
| ||
| SAMPLE FLOW | <1400 cm3/min |
| ||
| WARN |
| or | Dirty particulate filter |
|
|
| Leak downstream of critical flow orifice |
| ||
|
|
| 3 |
| |
|
| > 2200 cm /min. | Failed flow sensor/circuitry |
| |
|
|
|
|
| |
|
|
|
| If sample pressure is < 10 |
|
|
| Sample Pressure is | o Blocked particulate filter |
| |
|
| o Blocked sample inlet/gas line |
| ||
|
| <10 |
| ||
| SAMPLE PRES | o Failed pressure sensor/circuitry |
| ||
| Normally 29.92 |
| |||
| WARN | decreasing at 1 | If sample pressure is > 35 |
| |
| o Pressurized sample gas. Install vent |
| |||
|
| altitude (with no flow – pump |
| ||
|
| disconnected). | o Blocked vent line on pressurized sample/zero/span gas |
| |
|
|
|
| supply |
|
|
|
|
| o Bad pressure sensor/circuitry |
|
|
|
|
| Ambient temperature outside of specified range |
|
|
| Sample temperature is < 10oC or > | Failed bench heater |
| |
| SAMPLE TEMP | Failed bench temperature sensor |
| ||
| WARN |
| 100oC. | Relay controlling the bench heater |
|
|
|
|
| Failed relay board |
|
|
|
|
| I2C buss |
|
|
|
|
| GFC wheel stopped |
|
|
| Occurs when CO Ref is <1250 | Failed sync/demod board |
| |
|
| If status LED’s on the sync/demod board ARE flashing the |
| ||
|
| mVDC or >4950 mVDC. |
| ||
| SOURCE | cause is most likely a failed: |
| ||
|
|
|
| ||
| WARNING | Either of these conditions will result | IR source |
| |
|
| Relay board |
| ||
|
| in an invalid M/R ratio. |
| ||
|
| I2C buss |
| ||
|
|
|
|
| |
|
|
|
| IR |
|
|
|
|
| This message occurs at power on. If you have not cycled the |
|
|
| The computer has rebooted. | power on your instrument: |
| |
| SYSTEM RESET | o Failed +5 VDC power, |
| ||
|
|
|
| ||
|
|
|
| o Fatal error caused software to restart |
|
|
|
|
| o Loose connector/wiring |
|
|
|
|
| Blocked cooling vents below GFC Assembly. Make sure that |
|
|
|
|
| adequate clear space beneath the analyzer. |
|
|
|
|
| Analyzer’s top cover removed |
|
| WHEEL TEMP | The filter wheel temperature is | Wheel heater |
| |
| WARNING | controlled at 68 ± 2 °C | Wheel temperature sensor |
| |
|
|
|
| Relay controlling the wheel heater |
|
|
|
|
| Entire relay board |
|
|
|
|
| I2C buss |
|